Storage, Troubleshooting, If a calibration error occurs, refer to – Hach-Lange ISENa38103 User Manual

Page 12: If a measurement error occurs, refer to

Advertising
background image

Storage

For the best probe performance, do not let the reference junction dry out.
Short-term storage

1. Put the filling-hole cap in the filling hole (

Figure 1

on page 2).

2. Store the probe in a 25 mL of 100 mg/L Na+ standard solution with one Sodium Ionic

Strength Adjustor (ISA) powder pillow (0.4 g).

Long-term storage

1. Put the filling-hole cap in the filling hole (refer to

Figure 1

on page 2).

2. Rinse the probe with ISA rinse solution (refer to

Prepare the ISA rinse solution

on page 11). Never rinse with deionized water. Dry the probe with a lint-free cloth.

3. Fill the probe soaker bottle half full with 0.02 M NH

4

Cl Electrode Filling Solution..

4. Loosen the soaker bottle cap and put the soaker bottle on the probe.
5. Turn the soaker bottle cap clockwise to tighten the soaker bottle cap.
6. Make sure that the solution in the soaker bottle completely covers the glass bulb and

reference junction holes.

7. Before use after long-term storage, condition the probe in 25 mL of 100 mg/L Na+

standard solution with one Sodium ISA pillow (0.4 g) for at least 8 hours.

Troubleshooting

Message or symptom

Possible cause

Action

Probe not supported

Software not updated

To download the most current version of the
software, refer to the applicable product page on
the manufacturer's website.
Refer to the HQd Series meter manual for
specific instructions for the meter model.

HQd meter does not support
IntelliCAL

®

probe

Contact a Technical Support Representative.

Connect a probe or probe
requires service

Probe not connected properly

Disconnect, then connect the probe. Tighten the
locking nut.

Software not updated

To download the most current version of the
software, refer to the applicable product page on
the manufacturer's website.
Refer to the HQd Series meter manual.

Large number of methods stored on
probe

Continue to let probe connect. Do not disconnect
probe.

Damaged probe

Make sure that there is connectivity with another
probe or meter to confirm isolated issue with
probe. Contact a Technical Support
Representative.

mV reading is the same for all
solutions

Soaker bottle not removed

Remove the soaker bottle.

Electrical issue

Contact a Technical Support Representative.

Standard not recognized error

Soaker bottle not removed

Remove the soaker bottle.

Incorrect or contaminated standard
solution

Use fresh standard solution as specified in the
method.

Contaminated filling solution

Replace the filling solution.

12

Advertising