Change calibration options – Hach-Lange ISENa38103 User Manual

Page 9

Advertising
background image

Option

Description

Stability Criteria

When Auto Stabilization is off, sets the stability criteria—0.1 to
9.9 mV/min.

Lower stability criteria will require longer stabilization times, but the
measurement will be more precise.

Higher stability criteria will require shorter stabilization times, but the
measurements may be less precise.

Measurement
Limits

Sets the measurement limits—Lower limit (default: 0.023 mg/L) or Upper
limit (default: 14.00 mg/L).
The measurement limits can be set to match the acceptable values for
the sample. When the measurement is above the upper limit setting or
below the lower limit setting, the meter shows an "Out of limits" message.
This message is an alert to a potential problem with the process
conditions.

5. If prompted, enter a name for the new method settings. Additional changes made to

the settings of an existing method are automatically saved with the same method
name.

6. Push EXIT until the meter returns to the measurement mode.

Change calibration options

1. Make sure a probe is connected to the meter.
2. Push and select ISENa381 Settings.
3. Select Modify Current Settings.
4. Select Calibration Options and update the settings:

Option

Description

Std Set

Sets the temperature compensated standard sets that are used for
calibration—

10, 100 or 1000 mg/L

100 or 1000 mg/L

Standard set values are shown on the Calibration Options screen.
Custom standard sets are characterized at 25 °C (77 °F). Custom
standard values are not temperature compensated. Select the Custom
buffer to make a custom standard. Up to five standard values can be
made (refer to

Table 3

).

Note: Only the minimum calibration points must be measured for Done to

be shown on the calibration screen.

Chemical Form

Sets the chemical form.

Calibration
Units

Sets the preferred unit for ISE Calibration—mg/L (default), µg/L (available
only for custom calibration set), g/L, g/kg, mol/L, mmol/L, mol/kg, %, ppm
or ppb.

Std Set Values

When Std Set is set to Custom, sets the standard set values (refer to

Table 3

).

Up to five standard values can be made. Each standard value can include
a standard set value, Custom or No Standard.

Minimum Cal
Points

Sets the minimum number of calibration points necessary before a
calibration can be completed—2 or 3.

Slope Limit

Sets the slope limit—1 to 15% (acceptable slope criteria, default = 10%).
The slope must fall within set limits for successful calibration.

9