Change measurement options – Hach-Lange ISENa38103 User Manual

Page 8

Advertising
background image

The settings that can be changed are shown in

Table 2

.

Table 2 Parameter-specific settings

Setting

Options

Measurement Options

Units

Significant digits

Auto stabilization

Stability criteria

Upper and lower range limits

Calibration Options

Standard set

Calibration units

Minimum calibration points

Slope limit

Calibration reminder

Check Standard Options

Standard

Check standard reminder

Acceptance criteria

Change measurement options

Methods are groups of default or user-defined settings relevant to specific applications. If
the meter is set to the default method and the Modify Current Settings option is chosen, a
prompt for a new name is shown after the changes are entered. The settings are saved
with this name to distinguish them from the default method settings, which cannot be
changed. A saved method can be used instead of multiple adjustments to the individual
settings. Changes made to a user-defined method are automatically saved with the
existing name. Multiple methods can be saved for the same probe on each meter.

1. Make sure a probe is connected to the meter.
2. Push and select ISENa381 Settings.
3. Select Modify Current Settings.
4. Select Measurement Options and update the settings:

Option

Description

Units

Sets the preferred unit for ISE measurements—mg/L (default), µg/L, g/L,
g/kg, mol/L, mmol/L, mol/kg, %, ppm or ppb.
Note: The mV units are shown when the detailed display is selected.

Significant Digits Sets the significant digits shown—2, 3 (default) or 4.

Auto
Stabilization

Sets auto stabilization—on (default) or off.
The default stability drift rate is 1.0 mV/min.

8