Change calibration options, Maintenance – Hach-Lange POLYMETRON 9586 User Manual

Page 21

Advertising
background image

4. Select the option for the output signal during calibration:

Option

Description

ACTIVE

The instrument sends the current measured output value during the calibration procedure.

HOLD

The sensor output value is held at the current measured value during the calibration procedure.

TRANSFER A preset output value is sent during calibration. Refer to the controller user manual to change

the preset value.

5. With the sensor in the process sample, push enter. The measured value is shown. Wait for the

value to stabilize and push enter.

6. With a certified secondary verification instrument measure the concentration value of the sample.

To avoid impurities in the sample take the measurement before the sample enters the flow
chamber. Use the arrow keys to enter this value if different from the value displayed and push
enter.

7. Review the calibration result:

• PASS—the sensor is calibrated and the calibration factor is displayed.
• FAIL—the calibration is outside of accepted limits. Clean the sensor and retry. Refer to

Troubleshooting

on page 22 for more information.

8. If the calibration passed, push enter to continue.
9. If the option for operator ID is set to YES in the CAL OPTIONS menu, enter an operator ID. Refer

to

Change calibration options

on page 21.

10. On the NEW SENSOR screen, select whether the sensor is new:

Option Description

YES

The sensor was not calibrated previously with this controller. The days of operation and previous
calibration curves for the sensor are reset.

NO

The sensor was calibrated previously with this controller.

11. With the sensor still in the process push enter. The output signal returns to the active state and

the measured sample value is shown on the measure screen.

Note: If the output mode is set to hold or transfer, select the delay time when the outputs return to the active
state.

Change calibration options

The user can set a calibration reminder or include an operator ID with calibration data from this
menu.

1. Push the menu key and select SENSOR SETUP>CALIBRATE.
2. If the pass code is enabled in the security menu for the controller, enter the pass code.
3. Select CAL OPTIONS and push enter.
4. Use the arrow keys to select an option and push enter.

Option

Description

CAL REMINDER Sets a reminder for the next calibration in days, months or years—select the required

delay from the list.

OP ID ON CAL

Includes an operator ID with calibration data—YES or NO (default). The ID is entered
during the calibration.

Maintenance

D A N G E R

Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.

English 21