Specifications

Specifications are subject to change without notice.

Panel

Specification

Details

Dimensions

817 x 300 x 140 mm (32.2 x 11.8 x 5.5 in)

Weight

14.6 kg (32.15 lbs)

Sample flow rate

10—15 liters/hour

Working pressure

0.5—6 bar (7—87 psi)

Ambient temperature

0—60 °C (-32—140 °F)

Relative humidity

10—90%

Working electrode

Platinum

Counter electrode

Stainless steel

Reference

Ag / AgCl / KCl 0.1 M

Measuring cell

Acrylic

Sample conditioner

DIPA (diisopropylamine)

European standards

EN 61326 Class A for EMC; EN 601010-1 for low voltage safety

International standards

cETLus

Sensor

Specification

Details

Measuring range

0—500 ppb hydrazine; 0—100 ppb carbohydrazide

Sensitivity

< 0.2 ppb

Repeatability

1 ppb or ± 2% of measurement whichever is the greater

Detection limit

≤ 1 ppb

Response time

< 60 seconds

Sample temperature measurement range

5—45 °C (41—113 °F)

English 3