6. Select Acceptance Criteria and update the settings:

Option

Description

Acceptance Limits

Sets the tolerance limits for check standard—0.005 pH (default) to
1.000 pH.

Cal Expires on Failure Recalibration is necessary if check standard fails—Yes or No.

7. If prompted, enter a name for the new method settings. Other changes made to the

settings of an existing method are automatically saved with the same method name.

8. Push EXIT until the meter goes back to the measurement mode.

Maintenance

Clean the probe

Clean the probe when there is contamination on the sensor. Symptoms of contamination
are:

Readings are not accurate or consistent.

The stabilization time is slow.

A calibration error occurs.

Contamination is visible on the probe (e.g., dirt).

1. Rinse the probe with deionized water. Blot dry with a no-lint cloth.
2. Soak the glass bulb of the probe in the applicable cleaning agent. Refer to

Table 4

.

Note: Remove KCl salt build-up from the probe with warm deionized water.

3. Rinse or soak the probe for 1 minute in deionized water.
4. Soak the probe in pH 4 buffer for 20 minutes.
5. Rinse the probe with deionized water.
6. Blot dry with a no-lint cloth.

Table 4 Cleaning agent

Contaminant

Cleaning agent

Soak time

General contaminants

Electrode cleaning solution

12–16 hours

Mineral deposits

0.1 N HCl solution

10–15 minutes

Fats, grease and oils

Warm, mild detergent solution

2 hours (maximum)

Protein deposits

Pepsin cleaning solution

3 hours (maximum)

Inorganic buildup

EDTA solution

3 hours (maximum)

Clean the sleeve junction

Clean the sleeve junction when particles prevent from open or close the sleeve junction
correctly.

1. Make sure that the fill hole is closed and invert the probe. This prevents electrolyte

flow from the probe.

2. Slowly move the sleeve junction away from the glass bulb to open it.

Note: If the sleeve will not open, soak the electrode in warm water until the salt crystals

dissolve.

3. Rinse the area under the sleeve junction with deionized water to remove all particles

or salt crystallization.

4. Slowly move the sleeve junction to the glass bulb to close it.
5. Dry with a no-lint cloth.