Product overview, Preparation for use – Hach-Lange PHC745 User Manual

Page 2

Advertising
background image

Specifications

Details

Storage solution

Saturated KCl

Minimum sample depth

18 mm (0.71 in.)

Dimensions

Body diameter: 12 mm (0.47 in.)
Body length: 103 mm (4.1 in.)
Total length: 200 mm (7.87 in.)
Cable length: 1 m (3.28 ft)

Cable connection

M12 digital output and connector compatible with HQd meters

1

Radiometer Analytical’s Red Rod technology: Encapsulation of reference elements with a red protection tube.

Product overview

The probe is a digital pH combination electrode with refillable electrolyte, a glass body
and an integrated temperature sensor. The probe is for laboratory use and has a glass
sleeve junction that measures pH in wastewater, emulsions and viscous liquids.

Note: Do not use the probe with organic solvents or high alkaline sample types.

A 59-mL bottle of reference electrolyte filling solution (saturated KCl solution) is included
with the probe. Refer to

Figure 1

.

Figure 1 Probe overview

1 Storage protector

4 Cable, 1 meter (3.28 ft)

2 Filling-hole cap

5 Sleeve junction

3 Protective tape and filling-hole

6 Glass bulb and temperature sensor

Preparation for use

Prepare the probe before a calibration or a sample measurement.

1. Turn the storage protector clockwise to loosen it.
2. Remove the storage protector.
3. Fully rinse the reference junction and the glass bulb with deionized water to remove

all of the storage solution or possible salt particles. Dry with a no-lint cloth.

4. Remove the protective tape from the filling hole before initial use (refer to

Product

overview

on page 2). Discard the protective tape.

5. Add the filling solution to the probe as necessary (refer to

Fill the probe

on page 12). The filling solution must be above the standard solution or the sample
level during measurement or calibration (

Figure 2

).

2