6. If prompted, enter a name for the new method settings. Other changes made to the

settings of an existing method are automatically saved with the same method name.

7. Push EXIT until the meter goes back to the measurement mode.

Change calibration options

1. Make sure that a probe is connected to the meter.
2. Push and select PHC805 Settings.
3. Select Modify Current Settings.
4. Select Calibration Options and update the settings:

Option

Description

Buffer Set

Sets the temperature compensated buffer set for calibration:

Color Coded—4.01, 7.00, 10.01 (default)

IUPAC—4.01, 7.00, 10.01, 12.45

DIN—4.65, 9.23

IUPAC—4.01, 6.86, 10.01, 12.45

IUPAC—4.01, 6.86, 9.18, 12.45

IUPAC—4.01, 7.00, 9.18, 12.45

Custom Buffer Set (refer to

Table 3

)

Custom buffer sets are characterized at 25 °C (77 °F).
Buffer set values show on the Calibration Options screen.
Note: Measure only the minimum calibration points specified in Minimum

Cal Points to complete a calibration.

Buffer Set
Values

Sets the buffer values if Custom Buffer Set is selected (refer to

Table 3

).

Minimum Cal
Points

Sets the minimum number of calibration points that are necessary to
complete a calibration—1 (default), 2 or 3.

Slope Limit

Sets the slope limit—1% to 20% (acceptable slope criteria, default = 15%).
Set the slope between the set limits for a correct calibration.

5. Select the Calibration Reminder and update the settings:

Option

Description

Reminder
Repeat

The meter makes a sound when a calibration is necessary and at the
selected interval—Off (default), 2 h, 4 h, 8 h, 2 d, 5 d or 7 d.

Expires

Calibration expires after the selected time—Immediately, Reminder +
30 min, Reminder + 1 h, Reminder + 2 h or Continue Reading.
Note: The meter cannot read samples after calibration has expired unless

Continue Reading is selected.

6. If prompted, enter a name for the new method settings. Other changes made to the

settings of an existing method are automatically saved with the same method name.

7. Push EXIT until the meter goes back to the measurement mode.