Change measurement options, The settings that can be changed are shown in, Table 2 – Hach-Lange PHC745 User Manual

Page 8

Advertising
background image

Table 2 Parameter-specific settings

Setting

Options

Measurement Options

Resolution

Upper and lower range limits

Calibration Options

Buffer set

Buffer set values (if Custom Buffer Set selected)

Reminder

Minimum Cal points

Slope limit

Check Standard Options

Standard (temperature compensated buffer or custom at 25 °C (77 °F))

Reminder

Acceptance criteria

Standard value at 25 °C (77 °F) (if custom standard selected)

Units

pH

mV

Change measurement options

Methods are groups of default or user-defined settings related to specific applications. If
the meter is set to the default method, a prompt for a new name method shows. The
settings are saved with this name to identify them from the default method settings, which
cannot be changed. Use a saved method as an alternative of many adjustments to the
individual settings. Changes made to a user-defined method are automatically saved with
the existing name. Save different methods for the same probe on each meter.

1. Make sure that a probe is connected to the meter.
2. Push and select PHC805 Settings.
3. Select Modify Current Settings.
4. Select Units. Select pH (default) or mV.

Note: Use the mV option to find the probe offset in a pH 7 buffer or to measure the slope. Both

units show when the Detail Display mode is selected.

5. Select Measurement Options and update the settings:

Option

Description

Resolution

Set the resolution:

0.1 pH—Fast

0.01 pH—Fast (default)

0.01 pH—Medium

0.01 pH—Slow, or

0.001 pH—Slow

The resolution has an effect on the number of decimal digits and the
stabilization time. More stabilization time is necessary for higher resolution
measurements. Slower stabilization times give higher accuracy
measurements.

Measurement
Limits

Set the measurement limits—Lower limit (default: 0.00 pH) or Upper limit
(default: 14.00 pH).
Set the measurement limits to agree with the permitted values for the
sample. When the measurement is above the upper limit setting or below
the lower limit setting, the meter shows an "Out of limits" message. This
message is an alert to a possible problem with the process conditions.

8

Advertising