Two-point calibration – Hach-Lange SENSION+ DO6 User Manual

Page 17

Advertising
background image

13

6. Operation

ENGLISH

6.4.2. Two-point calibration

6.4.2.1. Preparation

1. Unscrew the sensor protector.
2. Dip the sensor into an oxygen-free solution.
3. Stir the sensor gently.

6.4.2.2. Calibration

Follow the diagram.

Note:
To perform a two-point calibration, you must be certain
that the solution is oxygen-free; otherwise, one-point
calibration (100% DO) using the calibrator tube is
preferable.
If calibration is not performed correctly, the membrane
must be cleaned and the electrolyte changed (see sen-
sor manual). The calibration is correct if an oxygen
content of over 98% is displayed when taking a mea-
surement in the calibrator tube. A value of over 99%
is optimal.
If the measuring instrument is not used for 5 minutes, it
shuts off automatically.

Dip the sen-
sor into an
oxygen-free
solution

sensION+ DO 6_HACH_LANGE_0710.indd ANG:13

sensION+ DO 6_HACH_LANGE_0710.indd ANG:13

14/9/10 10:53:34

14/9/10 10:53:34