Manual calibration, Preparation, Calibration – Hach-Lange SENSION+ DO6 User Manual

Page 18

Advertising
background image

14

6. Operation

6.5. Manual calibration

Important note: before turning on the instrument, check that the DO
sensor is connected and polarised.

Manual readjustment of the dissolved oxygen measured within the
range.
When readjusting DO, the instrument behaves as though it were per-
forming a one-point calibration.

6.5.1. Preparation

1. Unscrew the sensor protector.
2. Rinse the sensor with deionized water.
3. Dip the sensor into the solution in which manual calibration is to be

performed.

Note: Before beginning manual calibration, you must take a measure-
ment in the standard to be used in the calibration.

6.5.2. Calibration

Stir gently holding the sensor by the handle.
Follow the diagram.

(1)

Hold down

and then press

.

Note: If the measuring instrument is not used for 5 minutes, it shuts off
automatically.
If errors occur during calibration, an error message will appear on screen
(see pg. 19).
Before beginning manual calibration, you must take a measurement in the
standard to be used in the calibration.

Important note: to exit calibration without saving the adjustment,
press

.

(1)

sensION+ DO 6_HACH_LANGE_0710.indd ANG:14

sensION+ DO 6_HACH_LANGE_0710.indd ANG:14

14/9/10 10:53:34

14/9/10 10:53:34