Readjusting temperature, Operation english – Hach-Lange SENSION+ DO6 User Manual
Page 19
Advertising
15
6. Operation
ENGLISH
6.6. Readjusting temperature
The instrument can correct the deviation of a temperature
probe (built into DO sensors) at 25°C and 85°C (77°F
and 185°F).
This allows the instrument to be used as a precision ther-
mometer.
Follow the diagram.
2 s
Enter the correction of the
temperature sensor at
25 °C (77 °F).
Enter the correction of the
temperature sensor at
85 °C (185 °F).
To exit adjust-
ment without sav-
ing the data.
sensION+ DO 6_HACH_LANGE_0710.indd ANG:15
sensION+ DO 6_HACH_LANGE_0710.indd ANG:15
14/9/10 10:53:35
14/9/10 10:53:35