Set-up, Cell, Temperature coefficient (tc) – Hach-Lange SENSION+ EC5 User Manual
Page 13: Reference temperature (rt), Tds factor (f), English
9
5. Set-up
ENGLISH
The sensION
™
+
EC 5
is supplied configured with the following parameters:
- Cell: platinum (50 60).
- Temperature coefficient: 2.00%/°C.
- Reference temperature: 25°C.
- TDS calculation factor: 0.64.
To display and/or modify these param-
eters, follow the diagram below.
5.1. Cell
5060, Platinum cell: Pt.
5062, Titanium cell: tI.
Important note:
You must configure the correct cell to en-
sure correct measuring.
5.2. Temperature
coefficient (TC)
Enter the relevant coefficient value (0.00
to 9.99%/°C).
Recommended value for aqueous sam-
ples: 2.00%/°C.
5.2. Reference
temperature (RT)
Select a value between 20 and 25°C.
5.3. TDS factor (F)
Enter the relevant value
(0.00 to 4.44).