Technical specifications, English – Hach-Lange SENSION+ EC5 User Manual

Page 7

Advertising
background image

3

1. Technical Specifications

Subject to change without notice.

Specifications

Display

Liquid crystal, backlit, with pictograms

Keypad

Membrane, 5 keys

Measuring ranges

(resolution depends on range)

Conductivity

0.01 µS/cm to 500 mS/cm

Salinity

0.0 to 1999 mg/L NaCl

2.0 to 50.0 g/L NaCl

TDS

0 mg/L to 500 g/L

Temperature

–20.0 to 150.0 °C
–4.0 to 302.0 °F

Measuring error (± 1 digit)

Conductivity

≤ 0.5 %

Salinity

≤ 0.5 %

TDS

≤ 0.5 %

Temperature

≤ 0.2 °C (0.4 °F)

Reproducibility (± 1 digit)

Conductivity

± 0.1 %

Salinity

± 0.1 %

TDS

± 0.1 %

Temperature

± 0.1 °C (0.1 °F)

Temperature compensation

With built-in Pt1000 temperature sensor

Reference temperature (RT)

20 or 25 °C (68 or 77 °F). Factory settings: 25 °C (77 °F)

Temperature coefficient (TC)

0.00 to 5.00%/°C (factory settings: 2%/ °C.)

TDS conversion factor

0.00 to 4.44 (factory settings: 0.64)

Calibration standards

147 µS/cm, 1413 µS/cm, 12.88 mS/cm and 111.8 mS/cm (at 25 °C/77 °F)

Accepted cell constant

Between 0.05 and 50 cm

-1

(factory settings: 1.0 cm

-1

)

Dimensions of the measuring instrument and ambient conditions

Operating temperature

0 °C to 50 °C (32 °F to 122 °F)

Storage temperature

–15 °C to 65 °C (5 °F to 149 °F)

Ambient conditions

80% relative humidity (non-condensing)

Weight

300 g

Size

186 x 73 x 38 mm

Additional technical data

Energy management

Automatic power off after 5 minutes of inactivity

Type of protection

IP 67

Power supply (batteries)

Three 1.5V AA batteries. Autonomy of over 400 hours

Connector

MP-5, five-contact multipin

Warranty

Warranty

2 years

ENGLISH

sensION+ EC5 HACH_LANGE 0710.indd ANG:3

sensION+ EC5 HACH_LANGE 0710.indd ANG:3

6/10/10 12:03:02

6/10/10 12:03:02

Advertising