Readjusting temperature, Operation english – Hach-Lange SENSION+ EC5 User Manual
Page 17
Advertising
13
6. Operation
ENGLISH
6.4. Readjusting temperature
The instrument can correct the deviation of a temperature
probe (built into conductivity cells) at 25°C and 85°C (77°F
and 185°F).
This allows the instrument to be used as a precision thermom-
eter.
Follow the diagram.
Note: the process must be completed in order to store the
readjustment data.
2 s
μS/cm
@25ºC
μS/cm
ºC
EC
T
t
Enter the correction of
the temperature sensor at
25°C (77 °F).
To exit adjustment
without saving the
data
Enter the correction of
the temperature sensor at
85°C (185 °F).
sensION+ EC5 HACH_LANGE 0710.indd ANG:13
sensION+ EC5 HACH_LANGE 0710.indd ANG:13
6/10/10 12:03:25
6/10/10 12:03:25