Error messages, English – Hach-Lange SENSION+ EC5 User Manual
Page 23
Advertising
19
8. Error messages
Display
Caused by
Action
Measurement (EC, NaCl, TDS and temperature)
out of range.
Check sensor.
Low battery charge.
Replace the batteries.
E1
Unstable reading during a stability measurement
or during calibration.
100-second time-out.
Check that the sensor is properly immersed in
the sample.
E2
During calibration, the difference between the
constants obtained > 30%
Check standards.
Check sensor.
E3
During calibration,
the cell constant obtained < 0.05 cm
-1
Check standards.
Check sensor.
E4
Salinity cannot be calculated. TC = 0
Modify TC
E5
Same standards
Check standards.
Check sensor.
ENGLISH
sensION+ EC5 HACH_LANGE 0710.indd ANG:19
sensION+ EC5 HACH_LANGE 0710.indd ANG:19
6/10/10 12:03:30
6/10/10 12:03:30