Error messages, English – Hach-Lange SENSION+ EC5 User Manual

Page 23

Advertising
background image

19

8. Error messages

Display

Caused by

Action

Measurement (EC, NaCl, TDS and temperature)
out of range.

Check sensor.

Low battery charge.

Replace the batteries.

E1

Unstable reading during a stability measurement
or during calibration.
100-second time-out.

Check that the sensor is properly immersed in
the sample.

E2

During calibration, the difference between the
constants obtained > 30%

Check standards.
Check sensor.

E3

During calibration,
the cell constant obtained < 0.05 cm

-1

Check standards.
Check sensor.

E4

Salinity cannot be calculated. TC = 0

Modify TC

E5

Same standards

Check standards.
Check sensor.

ENGLISH