Preparation for use – Hach-Lange PHC10115 User Manual User Manual

Page 2

Advertising
background image

32.81, 49.21 or 98.42 ft) cable and is intended for field use. The probe provides pH
measurement in wastewater, drinking water and general aqueous applications. The probe
is not suitable for use with organic solvents or samples with pH less than 2.

Figure 1 Probe overview

1 Shroud (rugged model)

6 Locking ring (rugged model)

2 Probe storage cap

7 Rugged probe (5, 10, 15 or 30 meter cable)

3 Probe soaker bottle

8 Standard probe (1 or 3 meter cable)

4 Glass bulb and temperature sensor

9 Probe storage cap or soaker bottle holder

5 Reference junctions and protective tape

Preparation for use

To prepare the probe for initial use:

1. If a rugged probe, remove the shroud and probe storage cap (refer to

Remove the

shroud

on page 10).

2. If a standard probe, turn the probe soaker bottle cap counter-clockwise to loosen the

cap. Remove the soaker bottle from the probe.

3. Remove the protective tape from the reference junctions (refer to

Figure 1

on page 2). Dispose of the protective tape.

4. If the probe will be used immediately, prepare the probe for calibration or sample

measurement.

5. If the probe will not be used immediately, store the probe (refer to

Storage

on page 10).

To prepare the probe for calibration or sample measurement:

1. If a rugged probe, remove the probe storage cap.
2. If a standard probe, turn the probe soaker bottle cap counter-clockwise to loosen the

cap. Remove the soaker bottle from the probe.

3. Rinse the reference junctions and glass bulb thoroughly with deionized water to

remove the 3 M KCl solution completely. Blot dry with a lint-free cloth.

4. For the best stabilization time, condition the probe for several minutes in the sample

before use.

5. If a rugged probe, make sure that the shroud is installed before field use (refer to

Install the shroud

on page 10).

2

Advertising