6. Push Read.
Stir gently. The
display will show
"Stabilizing" and a
progress bar as
the probe
stabilizes in the
standard. The
display shows the
buffer that has just
been read and
shows the
temperature
corrected pH
value when the
reading is stable.

7. Repeat steps
4 - 6 until the
minimum number
of calibration
points specified in
the current
method have been
collected.

8. Push Done to
view the
calibration
summary. The
display will not
show Done until
the minimum
number of
calibration points
have been
collected.

9. Push Store to
accept the
calibration and go
back to
measurement
mode. If a rugged
probe, install the
shroud on the
probe (refer to

Install the shroud

on page 10).

Sample measurement

Before measurement:

The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.

If complete traceability is necessary, enter a sample ID and operator ID before measurement. Refer to the HQd meter
manual for more information.

Regular calibration is required for the best measurement accuracy (refer to

Calibration

on page 3).

Prepare the probe for use (refer to

Preparation for use

on page 2).

To deploy a rugged probe at a distance, toss the probe body with a gentle underhand throw. Do not swing the probe by the
cable as this may cause injury to the user, will cause severe strain on the cable and will shorten the service life of the probe.

Measurement notes:

Data is automatically stored in the data log when Press to Read or Interval is
selected in the Measurement Mode. When Continuous is selected, data will only be
stored when Store is selected.

Air bubbles under the sensor tip when submerged can cause slow response or error
in measurement. If bubbles are present, gently shake the probe until bubbles are
removed.

If a measurement error occurs, refer to

Troubleshooting

on page 10.

Measurement procedure:

4