Figure 6 pump adjustment, Sipper module – Hach-Lange Sipper Module for DR 5000 User Manual

Page 5

Advertising
background image

5

Sipper module

6.

Secure the Sipper Module and the Multi Cell Holder with the two locking screws
(

Figure 1 on page 2

).

7.

Contact with instrument has now been established.

SIPPER OPTIONS

is now available

in Instrument Setup.

8.

Pull the pump adjustment forward (item 1,

Figure 6 on page 5

) and open the pump

tubing clamp (item 2,

Figure 6

). Wrap the white pump tubing around the pump and

clamp the ends on the right and left in the two front retainers (item 3,

Figure 6

).The

push-in connectors of the pump tubing must be positioned as in

Figure 6

.

9.

Clean the Sipper Cell with a lint-free cloth and insert the Sipper Cell (item 1,

Figure 7

on page 6

).

10. Use the outlet connector (item 2,

Figure 7

) to join the white pump tubing

(item 3,

Figure 7

) to the Sipper Cell outlet (item 1,

Figure 7

).

11. Connect the drain tube (item 5,

Figure 7

) to the right end of the white pump tubing.

12. Connect the inlet tube to the inlet of the Sipper Cell (item 4,

Figure 7

).

Figure 6 Pump Adjustment

1

Pump Adjust Forward

2

Pump Tubing Clamp

3

Front Retainers