Sipper module

Performing a Single or Multi-Wavelength Measurement

1.

Load and insert the Sipper Module. The

SIPPER OPTIONS

key will appear on the

Instrument Setup screen.

2.

Press

SIPPER OPTIONS

to select the Sipper Settings. Select the sipper settings and

press OK.

3.

Press

OPTIONS

on the Single or Multi-Wavelength menu to configure the parameters.

4.

Route the drain tube to an appropriate drain or collection vessel.

5.

Place the Sample Inlet Tube into the blank and press

ZERO

. Leave the inlet tube in the

sample until the Sipper Pump stops and the settling cycle begins.

Note: The remaining time in seconds is shown instead of the date on the display. Press

CANCEL

to stop the sip cycle.

6.

When the settling cycle is complete, the blank is displayed. In the automatic mode, the
purge begins immediately after the reading without operator intervention. If the purge
is set up for a manual start, the instrument will wait for the operator to press

PURGE

.

Note: Distilled water or the sample can be used as the rinsing liquid.

7.

Rather than wasting the sample, a distilled water purge can be placed at the sample
inlet to rinse the sample cell between readings.