Calibration procedure, View the calibration data, Set the calibration reminder – Hach-Lange SENSION+ MM340 User Manual

Page 10: Sample measurements

Advertising
background image

Option

Description

Data introduction

Manual probe constant introduction.

Theoretical calibration

The probe calibration data is replaced at 25 °C
(77 °F).

Calibration procedure

This procedure is for general use with liquid calibration solutions. Refer
to the documents that are included with each probe for additional
information.

Note: Solutions should be stirred during calibration. For more information about the
stirring settings, refer to

Change the stirring settings

on page 12.

1. Pour the buffers or calibration solutions into the labeled calibration

beakers.

2. From the main menu use the or and and to select the

CALIBRATION parameter. Confirm.

3. If required select the Operator ID (1 to 10) and confirm.
4. Rinse the probe with deionized water and put the probe into the first

calibration beaker. Be sure that there are no air bubbles in the
membrane.

5. Push to Start calibration.
6. Push to measure the first calibration solution.

The next calibration solution is shown.

7. Rinse the probe with deionized water and put the probe into the

second calibration beaker. Be sure that there are no air bubbles in
the membrane.

8. Push to measure the second calibration solution.

The next calibration solution is shown.

9. Rinse the probe with deionized water and put the probe into the third

calibration beaker. Be sure that there are no air bubbles in the
membrane.

10. Push to measure the third calibration solution.

When the calibration is good, the display briefly shows Calibration
OK and then returns to the main menu.

Note: When a printer is connected the print menu opens and the result can be
printed.

View the calibration data

Data from the most recent calibration can be shown.

1. From the main menu use the or to select DATA LOGGER.

Confirm.

2. Select Display data.
3. Select Calibration data and confirm with . The last calibration data

is shown.

• pH-the slope and offset values are shown alternating with the

deviation (in %) and calibration temperature.

• ORP-the measured mV value and calibration temperature are

shown.

• Conductivity-the cell constant and calibration temperature for each

standard are shown.

Set the calibration reminder

The calibration reminder can be set between 0 to 23 hours or 1-7 days
(default 1 day). The display shows the remaining time to the new
calibration.

Note: When 0 days is selected, the calibration reminder is turned off.

1. From the main menu use the or to select CALIBRATION.

Confirm.

2. Use the to enter the calibration menu.
3. Use the or to select Cal. frequency and confirm.
4. Use the and to advance to the next step and use the or to

change a value. Confirm.
Push to start the calibration.

Sample measurements

Each probe has specific preparation steps and procedures for taking
sample measurements.

1. From the main menu use the or and and to select

MEASURE. Confirm.

10 English