Soluzioni tampone (DIN 19266)
Fare riferimento ai valori di pH e ORP (mV) in
soluzioni tampone specifiche a diverse temperature.
Tabella 4 Valori di pH, ORP (mV) e temperatura
Temperatura
pH
mV
°C
°F
5
32
1,668
4,004
6,951
7,087
9,395
10,245
13,207
10
50
1,670
4,000
6,923
7,059
9,332
10,179
13,003
20
68
1,675
4,001
6,881
7,016
9,225
10,062
12,627
25
77
1,679
4,006
6,865
7,000
9,180
10,012 12,454
30
86
1,683
4,012
6,853
6,987
9,139
9,966
12,289
40
104
1,694
4,031
6,838
6,970
9,068
9,889
11,984
50
122
1,707
4,057
6,833
6,964
9,011
9,828
11,705
60
140
1,723
4,085
6,836
6,968
8,962
—
11,449
70
158
1,743
4,126
6,845
6,982
8,921
—
—
80
176
1,766
4,164
6,859
7,004
8,885
—
—
90
194
1,792
4,205
6,877
7,034
8,850
—
—
Italiano 55