Change the language

The display language is selected when the meter is powered on for the
first time.

1. Use the or to select a language from a list.
2. Confirm with . The measurement screen shows DATA OUTPUT.
3. Select Deactivated if no printer or PC is connected and confirm.

Refer to

Select the data output

on page 13 for more information

about Data Output.

Standard operation

Setup for dissolved oxygen

Before a calibration or a measurement for dissolved oxygen, the probe
must be polarized and the settings for atmospheric pressure and salinity
must be entered.

Polarize the electrode

If the probe or batteries are disconnected, connect the probe or install
the batteries and wait for polarization:

Disconnect time

Polarization time

< 5 minutes

10 minutes

5 to 15 minutes

45 minutes

> 15 minutes

6 hours

Calibration

W A R N I N G

Chemical exposure hazard. Obey laboratory safety procedures and wear all of
the personal protective equipment appropriate to the chemicals that are handled.
Refer to the current material safety data sheets (MSDS) for safety protocols.

Calibration settings

The calibration settings contain Calibration type, Calibration frequency
and Display options.

1. From the main menu use the or to select CALIBRATION.

Confirm.

2. Use the to enter the calibration menu.
3. Use the or to select the following options:

Option

Description

Stability C.:

Criterion by stability—select Fast, Standard or Strict.

Calibration
type

pH: Calibration type—select Technical buffers,
DIN19266 Buffers, User Buffers, To a X value, data
introduction or Theoretical calibration.EC: Calibration type
—select Molar Standards, Demal Standards, NaCl
Standards, Calibration to a X value, data introduction or
Theoretical calibration. Refer to

Calibration types

for more

information.DO: Calibration type—select Water saturated
air, Water saturated air & Zero, To a X value or Theoretical
calibration.

Cal. frequency

Calibration reminder—can be set between 0-23 hours (pH)
or 0–7 days (EC). The display shows the remaining time to
the new calibration. Refer to

Set the calibration reminder

on page 12 reminder for more information.

Display mV

Display mV—select YES or NO to show mV.

Display nA

Display nA—select YES or NO to show DO current in nA
(nanoampere).

Calibration types
Different calibration types can be selected.

1. From the main menu use the or to select CALIBRATION.

Confirm.

2. Use the to enter the calibration menu.
3. Use the or to select Calibration type.

Option

Description

Technical buffers

pH 2.00, 4.01, 7.00, 9.21 and 10.01 at 25 °C
(77 °F)