Table 2 Measurement warnings and errors

Error/Warning

Solution

- - - - -

Measured value is out of range.
Examine the probe: Clean the probe (refer to

Clean

the probe

on page 16 for more information); make sure

that there are no air bubbles in the membrane. Shake
the probe like a thermometer; connect a different
probe to find if the problem is with the probe or the
meter.

Unstable measurement
(pH, EC and DO
measurements)
Time > 120 s

Make sure that the membrane and the diaphragm are
properly immersed in the sample.
Examine the temperature.
Examine the probe: Clean the probe (refer to

Clean

the probe

on page 16 for more information); make sure

that there are no air bubbles in the membrane. Shake
the probe like a thermometer; connect a different
probe to make sure that if problem is with probe or
meter.

Unstable measurement
(ISE measurements)
Time > 240 s

Make sure that the membrane and the diaphragm are
properly immersed in the sample.
Examine the temperature.
Examine the probe: Clean the probe (refer to

Clean

the probe

on page 16 for more information); make sure

that there are no air bubbles in the membrane. Shake
the probe like a thermometer; connect a different
probe to make sure that if problem is with probe or
meter.

TC = 0
Impossible to measure
salinity

Modify TC

10.389 mg/L >>>>>
00012 00:13

ISE measurement: The measured value is superior to
the highest standard used in calibration. Measure
again.

0.886 mg/L <<<<<<
00018 00:11

ISE measurement: The measured value is inferior to
the lowest standard used in calibration. Measure
again.

Replacement parts and accessories

Note: Product and Article numbers may vary for some selling regions. Contact the
appropriate distributor or refer to the company website for contact information.

Replacement parts

Description

Item no.

sensION+ PH3 Lab pH-meter with accessories, without
probe

LPV2000.98.0002

sensION+ PH31 Lab pH-meter, GLP, with accessories,
without probe

LPV2100.98.0002

sensION+ MM340 Lab pH & Ion-meter, GLP, 2 channels,
with accessories, without probe

LPV2200.98.0002

sensION+ EC7 Lab conductivity meter, with accessories,
without probe

LPV3010.98.0002

sensION+ EC71 Lab conductivity meter, GLP, with
accessories, without probe

LPV3110.98.0002

sensION+ MM374, 2 channel Lab meter, GLP,
accessories, without probes

LPV4110.98.0002

sensION+ MM378, 2 channel Lab meter, GLP,
accessories, without probes

LPV4130.98.0002

Consumables

Description

Item no.

pH buffer solution 4.01, 250 mL

LZW9463.99

pH buffer solution 7.00, 250 mL

LZW9464.98

pH buffer solution 10.01, 250 mL

LZW9471.99

Electrolytic solution (KCl 3M), 125 mL

LZW9510.99

Electrolytic solution (KCl 3M), 250 mL

LZW9500.99

Electrolytic solution (KCl 3M), 50 mL

LZW9509.99

English 19

Advertising