Change the stirring settings, Turn the stirrer on or off, Change the temperature units – Hach-Lange SENSION+ MM378 User Manual

Page 14: Data logger, Display data, Stirring settings, refer to

Advertising
background image

Example: reference thermometer: 24.5 °C; meter: 24.3 °C.
Adjustment value: 0.2 °C.

3. Enter the adjustment value for the 25 °C reading:

a. From the main menu use the or to select SYSTEM. Confirm.
b. Use the or to select Readjust temp. and confirm.
c. Select Channel 1 or Channel 2 and confirm.
d. Use the or to select 25 °C and confirm.
e. Use the arrow keys to enter the adjustment value for 25 °C.

Confirm.

Option

Description

Channel 1 Select Parameter 1 or Parameter 2 . Parameter 1 the

temperature is measured by the pH probe. Parameter 2 the
temperature is measured by the conductivity cell.
If one of the parameters is not used, the meter automatically
assigns the temperature of the other parameter. If no A.T.C. is
detected, the temperature value must be entered manually

Channel 2 The temperature for Channel 2 is measured by the DO probe or

manually entered.

4. Put the probe and a reference thermometer in a container of water at

approximately 85 °C and allow the temperature to stabilize.

5. Compare the temperature from the meter with that of the reference

thermometer. The difference is the adjustment value for the meter.

a. Use the or to select 85 °C and confirm.
b. Use the arrow keys to enter the adjustment value for 85 °C.

Confirm.

c. Select Save changes and confirm.

Change the stirring settings

The stirring speed can be changed during calibration and during a
measurement.

1. Use the or to change the stirring speed during calibration and

during a measurement.

Turn the stirrer on or off

The stirrer 1 works with Channel 1 and 2 (Stirrer 1). A second stirrer can
be connected to channel 2 (Stirrer 2). For the stirrer 2 activation refer to
the following steps.

1. From the main menu use the or to select SYSTEM. Confirm

with .

2. Use the or to select Stirrer N.2 and confirm with .
3. Use the or to select YES to turn the stirrer 2 on.

Note: Select NO to turn the stirrer 2 off.

Change the temperature units

The temperature units can be changed to Celsius or Fahrenheit.

1. From the main menu use the or to select SYSTEM. Confirm.
2. Use the or to select Temperature units and confirm.
3. Use the or to select between Celsius or Fahrenheit and confirm.

Data logger

Display data

The Display data log contains Measurement data, Electrode report and
Calibration data. The stored data can be send to a printer or to a PC.
When the data log becomes full (400 data points), the oldest data point
is deleted when a new data point is added.

1. Use the and to select Channel 1 or Channel 2 and confirm.
2. From the main menu use the or to select DATA LOGGER and

confirm.

3. Use the or to select Display data and confirm.
4. Use the or to select

Option

Description

Measurement data Measurement data—stores automatically each time a

sample is measured