Advanced Operations

6.8.2 Upgrade of the instrument software

To obtain the software for the update from the Internet at
www.hach-lange.com.

1. Go to http://www.hach-lange.com

2. Select the country and go to Download>Software.

3. Enter DR 5000 in ”Search for documents“.

4. Locate the appropriate download and follow the prompts for

saving the file(s) to the USB memory stick (

Section 9 on page

133

) or to the PC.

5. Unpack the ZIP file and save the files to the USB memory stick.

6. Press Instrument Update in the System Checks menu.

7. Connect the USB memory stick to the USB interface (type A)

on the instrument, see

section 3.3 on page 14

.

8. Press OK. The link is established automatically and the

software is updated.

9. Press OK to return to the System Checks menu.

Note: When the instrument software has been updated, a prompt to restart
the instrument is displayed.

6.8.3 Optical checks

For each optical check, the measured results must be evaluated
against user-specific requirements. The check options do not
define tolerances.

1. Press Optical Checks in the System Checks menu.

The Optical Checks menu contains programs for checking the
wavelength accuracy, stray light and photometric accuracy.

An optional test filter set (Verification Kit) (

Section 9 on page 133

)

containing six precision glass filters, target values, tolerances and
instructions is available as an aid for carrying out comprehensive
in-house instrument checks.