60

Standard Operations

5.5.6.3 Program a new test

1. The instrument needs to be at the Main Menu.

2. Insert the sample cuvette/vial or blank (depending on the

working procedure) into the cell compartment #1 and close the
cell compartment.

3. After "Program not available", a data overview is displayed,

including the specifications of the test that is to be revised.

The working procedures of the corresponding test contains the
new data (Wavelength, Factors, Measurement Ranges,
Conversion Factor for chemical Form etc.).

4. Highlight the corresponding line including the specifications of

the test that is to be programmed and press Edit.

5. To finish programming, press Store and then Cancel to return

to the Main Menu.

6. Press Start to run the test with the new data.

5.5.7 Upgrade of the instrument software

To obtain the software for the update from the Internet at
www.hach-lange.com:

1. Go to http://www.hach-lange.com

2. Select the country and go to Download>Software.

3. Enter DR 5000 in ”Search for documents“.

4. Locate the appropriate download and follow the prompts for

saving the file(s) to the USB memory stick (

Section 9 on page

133

) or to the PC.

5. Unpack the ZIP file and save the files to the USB memory stick.

6. Press Instrument Update in the System Checks menu.

7. Connect the USB memory stick to the USB interface (type A)

on the instrument, see

section 3.3 on page 14

.