Formazin calibration procedure – Hach-Lange 2100 N Laboratory Turbidimeter User Manual User Manual

Page 35

Advertising
background image

• In Calibration mode, automatic range and signal averaging on are

selected. When calibration is completed, all operational modes go
back to the last settings.

• All nephelometric (turbidity units of measure) calibrations are done at

the same time.

• Ratio-on and Ratio-off calibration data is measured and recorded at

the same time.

• Clean the USEPA filter assembly before doing a primary calibration,

or at least every 3 months (which is the USEPA-recommended
primary calibration interval).

Formazin calibration procedure

For the best accuracy, use four matched sample cells or the same sample cell for all measurements during calibration. Refer to

Matching sample cells

on page 20.

1. Remove the filter
assembly. Refer to

Change the filter
assembly

on page 36.

2. Clean and inspect
the lens of the USEPA
filter assembly. Refer to

Clean the filter
assembly

on page 36.

3. Hold the tab of the
USEPA filter assembly
so that the arrows point
toward the front of the
instrument. Push the
filter assembly fully into
the housing.

4. Push CAL.
The S0 light turns on.
The NTU value of the
dilution water used in
the last calibration is
shown.

5. Rinse a clean
sample cell two times
with dilution water. Fill
the sample cell to the
line (about 30 mL) with
dilution water and
immediately put the cap
on the sample cell. Use
the same dilution water
that was used to
prepare the formazin
standards.

6. Clean the sample
cell with a soft, lint-free
cloth to remove water
spots and fingerprints.
Do not invert the
sample cell.

English 33