Replace a fuse

D A N G E R

Fire hazard. Use the same type and current rating to replace fuses.

Replacement parts:

• Fuse for 115 V operation, time-delay, 250 V, 1.6 A (3030700), or
• Fuse for 230 V operation, time-delay, 250 V, 1.6 A (3030600)

To replace a fuse, refer to the illustrated steps in

Figure 11

.

Figure 11 Replace a fuse

Troubleshooting

Refer to the tables in this section for error codes, diagnostic codes,
common problem messages or symptoms, possible causes and
corrective actions.

Error codes

Table 7

lists the error codes shown for different conditions. Error codes

identify instrument malfunction or operator error.
The instrument continues operation in an error condition.
Push ENTER to clear an error code from the display.

Note: Any calibration being calculated when an error occurs, is discarded. The old
calibration is kept.

Table 7 Error codes

Error

Description

Solution

ERR01 The turbidity of the

dilution water is
greater than 0.5 NTU.

Start the calibration again with lower turbidity
dilution water.

Note: Ignore ERR01 when the sample cell diameter is
less than 25 mm. Push UNITS/Exit to go back to
measurement mode.

ERR02

• Two calibration

standards have the
same value.

• The difference

between two
calibration
standards is less
than 60.0 NTU.

• The turbidity of

Standard 1 is too
low (less than
10 NTU).

1. Inspect the preparation of standards.
2. Do the calibration again.

Note: Ignore ERR02 when the sample cell diameter is
less than 25 mm. Push UNITS/Exit to go back to
measurement mode.

38 English