Specifications

Specifications are subject to change without notice.

Specification

Details

Measurement method

Nephelometric

Regulatory

Meets EPA Method 180.1
ASTM D7315 - Standard Test Method for
Determination of Turbidity Above 1 Turbidity Unit (TU)
in Static Mode
ASTM D6655 - Standard Test Method for
Determination of Turbidity Below 5 NTU in Static Mode

Light source

Tungsten filament lamp

Measurement modes

NTU, NEPH (Nephelo) and EBC

Range

NTU (Ratio on, manual range): 0–0.999, 0–9.99, 0–
99.9, 0–4000
NTU (Ratio on, auto range): 0–4000 auto decimal
NTU (Ratio off): 0–40.0
Nephelo (Ratio on, manual range): 0–9.99, 0–99.9, 0–
26,800
Nephelo (Ratio on, auto range): 0–26,800 auto
decimal
Nephelo (Ratio off): 0–268
EBC (Ratio on, manual range): 0–0.999, 0–9.99, 0–
99.9, 0–980
EBC (Ratio on, auto range): 0–980 auto decimal
EBC (Ratio off): 0–9.8

Accuracy

1,

2

Ratio on: ±2% of reading plus 0.01 NTU from 0–
1000 NTU, ±5% of reading from 1000–4000 NTU
based on formazin primary standard
Ratio off: ±2% of reading plus 0.01 NTU from 0–
40 NTU (under reference conditions

3

)

Resolution

Turbidity: 0.001 NTU/EBC, Nephelo (on lowest range)

Specification

Details

Repeatability

±1% of reading or 0.01 NTU, whichever is greater
(under reference conditions)

Response time

Signal averaging off: 6.8 seconds
Signal averaging on: 14 seconds (when
10 measurements are used to calculate the average)

Stabilization time

Ratio on: 30 minutes after start-up
Ratio off: 60 minutes after start-up

Reading modes

Manual or auto range, signal averaging on or off, Ratio
on or off

Power requirement

115–230 VAC, 50/60 Hz (automatic power selection)
28 W maximum

Pollution
degree/installation
category

2; II

Protection Class

1

Operating conditions

Temperature: 0 to 40 °C (32 to 104 °F)
Relative humidity: 0–90% at 25 °C, 0–75% at 40 °C,
noncondensing
Altitude: 2000 m (6560 ft) maximum
Indoor use only

Storage conditions

–40 to 60 °C (–40 to 140 °F), instrument only

Interface

RS232C serial interface by way of DB9 subminiature
D-shell connector for data output to computer or
printer, and data input (command). No handshaking.

Air purge

Dry nitrogen or instrument grade air (ANSI MC 11.1,
1975)
0.1 scfm at 69 kPa (10 psig); 138 kPa (20 psig)
maximum
Hose barb connection for

1

/

8

-inch tubing

English 5