Flashing 9s, Flashing 0s, Replacement parts and accessories – Hach-Lange 2100 N Laboratory Turbidimeter User Manual User Manual

Page 42: Flashing 9s flashing 0s, Filter or use a glass-fiber filter. refer to, Accessories, Replacement parts and accessories (continued)

Advertising
background image

1. Turn off the instrument.
2. Push and hold CAL.
3. Turn on the instrument.

The CAL? light flashes. The instrument starts in Calibration mode.

4. Calibrate the instrument before use.

Flashing 9s

When manual ranging is selected, the display will flash all 9s when the
sample being measured is greater than the selected range.
When automatic ranging is selected, the display will flash 9s when the
sample is greater than the maximum range of the instrument. The
display will also flash 9s if Ratio is off and the measurement is greater
than 40 NTUs (268 nephelos or 9.8 EBCs). Turn Ratio on. Refer to

Measure over-range samples

on page 23.

Flashing 0s

When manual ranging is selected, the display will flash all 0s when the
sample measured is less than the selected range.
When automatic ranging is selected, the display will flash all 0s when the
measurement is less than the range of the instrument or a negative
value. Calibrate the instrument.

Replacement parts and accessories

Note: Product and Article numbers may vary for some selling regions. Contact the
appropriate distributor or refer to the company website for contact information.

Replacement parts

Description

Quantity

Item no.

Calibration kit, StablCal

®

, sealed sample

cells (<0.1, 20, 200, 1000 and 4000 NTU)

1

2662105

Cover, sample cell compartment

1

4702500

Cover, lamp access

1

4703200

Dust cover

1

4703000

Replacement parts and accessories (continued)

Description

Quantity

Item no.

Filter assembly, USEPA

1

3031200

Fuse for 115 V operation, time-delay,
250 V, 1.6 A, UL/CSA approved

1

3030700

Fuse for 230 V operation, time-delay,
250 V, 1.6 A, IEC type, VDE approved

1

3030600

Gelex

®

secondary turbidity standardization

kit
(stray light standard and 0–2, 0–20, 0–
200 and 200–4000 NTU)

1

2589000

Lamp replacement kit

1

4708900

Oiling cloth

1

4707600

Power cord, North America, 115 VAC,
UL/CSA approved

1

1801000

Power cord, European, 230 VAC, VDE
approved

1

4683600

Sample cells, 30mL, 1 in. round glass

6

2084900

Silicone oil

1

126936

Accessories

Description

Quantity

Item no.

Calibration kit, StablCal

®

, 500 mL each

(<0.1, 20, 200, 1000 and 4000 NTU)

1

2662110

Calibration kit, StablCal

®

, 500 mL each

(<0.1, 20, 200, 1000 and 4000 NTU)

1

2662100

Cable, computer, DB-9 to DB-9

1

4950200

Cell adapter, 12–13 mm

1

3033400

Cell adapter, 16 mm

1

3033500

40 English

Advertising