Instrument Use

Table 3 Stored Programs

Program

Description

COD Program

Heats vials for 2 hours at 150 °C. In the cooling phase, 4 beeps
indicates that the vials have cooled to 120 °C. Remove and
carefully invert the vials several times, before allowing them to
cool further in a rack.

TOC Program

Heats the vials for 2 hours at 105 °C. This setting is suitable for
all Hach TOC vial tests.

100 °C Program


Heats vials for 30, 60, or 120 minutes at 100 °C. For example,
the “100 °C, 60 minutes” setting is needed to digest samples
using the Methods Prep Set.

105 °C Program


Heats vials for 30, 60, or 120 minutes at 105 °C.

150 °C Program


Heats vials for 30, 60, or 120 minutes at 150 °C.

Select program
COD

TOC

COD:

150°C

120’

OK

Select program
COD

TOC

TOC:

105°C

120’

OK

Select program
100°

105°

Digestion period

100°C: 100°C

60’

OK

120’

60’

0’

3

Select program
100°

105°

Digestion period

105°C: 105°C

60’

OK

120’

60’

0’

3

Select program

150°

165°

Digestion period

150°C: 150°C

60’

OK

120’

60’

0’

3