2.5 Setting User Programs

The DRB 200 can accommodate three user programs in
storage locations PRG1, PRG2, and PRG3.

1.

Select PRG1 from the Select Program menu.

2.

Press Prog to enter the programming mode.

3.

Assign a 4-character program name using the
appropriate keys. Press OK to confirm the entry.

4.

Use the up and down arrow keys to set the
temperature value between 37–165 °C. Press OK to
confirm the entry.

5.

Use the up and down arrow keys to set the time
between 0-480 minutes. Press OK to confirm the entry.

6.

Press OK to confirm the program. The program can be
changed by pressing Prog. Press the up arrow key to
return to the program selection screen.

2.6 Error Messages

165 °C Program


Heats vials for 30, 60, or 120 minutes at 165 °C.

Table 3 Stored Programs (Continued)

Program

Description

Select program

150°

165°

Digestion period

120’

60’

10’

OK

165°C: 165°C

60’

PRG1: 100°C

60’

OK

Prog

Temperature?

Time?

60’

OK

Temperature?

115°C

OK

Program name?

Program name?

Select program
PRG1

PRG2

115 :

115°C

60’

100°C

OK

OK

( RG1)

P

5

(11

)

OK

OK

Prog

Error Message

Problem

“Block is too hot,
Please Wait”

The temperature of the heating block is higher than the
target temperature. Wait until the heating block cools.

“Init Error”

The instrument is defective. Contact Customer Service.

Advertising