Calibration setup, Multi-point calibration proving - continued – Liquid Controls LCR 600 Setup & Op User Manual
Page 31

31
calIbratIon Setup
multi-point calibration proving - continued
LCR-600: 60011
Adjacent linearization points out of
0.25% range.
The LCR 600 will not allow you to enter the Applied
setting of the Linearization Mode: until all adjacent
calibration points are within 0.25% of each other.
to prove a meter system to a product with multi-
point calibration:
10. Repeat steps 4 through 10 using a new flow rate for the
next Point #: until all the points between the lowest
flow rate and the maximum flow rate are within 0.25% of
each other.
11. Navigate to Calibration Setup 1. Move the pointer
to
Linearization Mode:
and select Applied.
The linearization points on Calibration
Setup 3 will automatically sort by flow rate
after the Applied
setting is selected.