Electronic Subsystem

The electronics subsystem provides the power for operation and control of the MAGPIX
system and communication between its parts.

Power Input Module

The power input module contains the input power plug, hard power toggle switch, and fuses.
This is the protective earthing point for the MAGPIX system. The mating power cord
connector type is IEC-320-C13. The mating power cord provides electrical power to the
instrument when it is connected to an electrical outlet and is the means of disconnection. The
power input auto-senses the voltage range.

FIGURE 39.

Power Input Module

1. Hard power toggle switch

2. Input power plug

WARNING:

Do not obstruct this means of disconnection. Connect only to
outlets that contain protective earthing. Before changing a fuse,
turn off the instrument and unplug the power cord to avoid any
danger of electrical shock.

Communications Port

The communications port connects MAGPIX to the computer. It is a USB port, labeled P1.