Annual procedures, Replacing the sample probe tube – Luminex MAGPIX Hardware and Installation Manual User Manual
Page 79
Advertising

10. Screw the syringe back into its housing.
FIGURE 56.
The Syringe
1. Syringe housing
2. Metal rod plunger
3. Syringe seal, containing black O-ring
4. Glass cylinder
5. Plunger guide
11. Return the plunger guide to its original position. The bottom of the plunger fits into the
indentation in the plunger guide.
12. Plug in the power cord and turn on MAGPIX.
13. Use the software to run the prime command twice, watching for any leaks in the syringe
area.
14. Close the side access door.
Annual Procedures
Replacing the Sample Probe Tube
The tube that connects the sample probe to the valve is subject to wear because of the
constant motion of the sample probe. Consequently, maintenance requires replacing it on a
yearly basis.
MAGPIX
®
66