Luminex 100 User Manual Version 1.7 Addendum

PN 89-00002-00-143 Rev. B

5

contacting Technical Support. You can move the line temporarily for
cleaning and maintenance purposes.

Sheath Delivery

System Setup

When adjusting the Luminex 100 regulator, turn it clockwise as far
as it will go to the fully closed position, then turn it back 1/2 of a
turn. Be careful not to force the regulator past the fully closed
position.

Do not move the Sheath Delivery (SD) system or the sheath line
while the system is in use.

Replacing the

Sheath Delivery

System with Sheath

Bottles

You may need to replace the Sheath Delivery (SD) system with
sheath bottles for maintenance or troubleshooting, as directed by
Luminex Technical Support.

‹

To replace the SD system with sheath bottles:

1. Power off and disconnect the SD system from the Luminex 100.

2. Turn the black regulator screw inside the center door of the

Luminex 100 one full turn counter-clockwise.

3. Attach the sheath and waste bottle containers. Make sure there is

sheath fluid in the bottle (to the fill line) and the sheath cap is
tightened.

4. Prime the system.

5. As the system is priming, adjust the black regulator screw inside

the center door of the Luminex 100. Turn clockwise to increase
or counterclockwise to decrease, until the sheath pressure value
is within +/- 0.1 psi of the normal level of your instrument.

6. When the prime function is complete, loosen the sheath bottle

cap to release pressure.

7. Tighten sheath cap and prime the system once again.

8. Make sure that the sheath and air pressure values increase to the

normal level once again.

Do not move the sheath bottle or sheath line while the system is in
use.

Probe Height

Adjustment

This information supplements the information in the LDS 1.7 User
Manual.