Running Calibration and Verification

Before running Calibration/Verification from the Auto Maint tab, you need to import Cal and
Ver kit information. Perform that procedure from the Lot Management tab. See

Importing

CAL or VER Kits

.

Run the Calibration/Performance Verification routine as part of your weekly maintenance
routine:

1. On the Home page, click System Initialization under Daily Activities. The Auto Maint

tab on the Maintenance page opens.

2. Click the Calibration Verification button under Automated Maintenance options.

3. Add the appropriate reagents to the off-plate reservoir and the strip well, using the

diagram in the Reagents pane of the Auto Maint tab to guide you.

4. Click Run.

Running the Performance Verification Routine

Run the Performance Verification routine as part of your daily startup routine.

1. On the Home page, click System Initialization under Daily Activities. The Auto Maint

opens.

2. On the Auto Maint tab, click Performance Verification.

3. Add the appropriate reagents to the off-plate reservoirs and the strip well, using the

diagram in the Reagents pane of the Auto Maint tab to guide you.

4. Click Run.

Lot Management Tab

Use this tab to manage calibration and verification kits.

Maintenance > Lot Management

Maintenance Page

95