Plate layout subtab – Luminex xPONENT 4.2 for FLEXMAP User Manual

Page 57

Advertising
background image

• 1:2 - Halves the standard from each previous iteration.

• 1:10 (Log) - Computes a value of one-tenth of the standard from each previous

iteration.

• 1/2 Log - Creates a 1:3.16 dilution, or half of each 1:10 (Log) from each previous

iteration.

• Apply Dilution - Applies the dilution selected in the Dilution list.

NOTE: The Dilution list and Apply Dilution button are displayed only if a

quantitative analysis has been selected.

NOTE: You can also manually type a number to set your own dilution factor.

It must be a whole number.

• Assay Control Information - Lists the selected control reagents. The list displays the

Reagent, Name, Lot Number, Expiration, and Manufacturer. Existing control lot
information can be applied or new information can be typed manually.

• Apply Ctrl Lot - Opens the Select Lot dialog box. Select a lot from the list and then

click OK.

• Show Concentration - Expected, Low, and High set the expected, lowest, or highest

acceptable concentration of the analyte in the sample.

• Apply Values - Applies a value down or across the list of analytes.

• Cancel - Returns to the Batches tab.

• Back - Returns to the previous tab.

• Next - Continues to the Plate Layout tab.

Plate Layout Subtab

Batches > Batches > Plate Layout

Batches Page

43