Revive after storage routine – Luminex xPONENT for MAGPIX IVD User Manual

Page 10

Advertising
background image

CAUTION:

Correct sample probe height is critical to successful sample
acquisition and calibration. Problems with the sample probe can
lead to fluid leaks and inhibit sample acquisition. To adjust the
height of the sample probe, follow the steps above.

FIGURE 1.

Sample Probe Height Adjustment

1. Well D6

2. Strip well SD1

3. Reservoir RB1

Revive After Storage Routine

NOTE: The Revive After Storage routine is necessary when the system

runs for the first time and is recommended when the system has
been idle for more than a week.

After you have adjusted the sample probe height, run the Revive After Storage (Luminex)
routine.

MAGPIX

®

User Quick Guide

4

For In Vitro Diagnostic Use