Creating calibration and verification reports – Luminex xPONENT for MAGPIX IVD User Manual

Page 13

Advertising
background image

1. Load the CAL/VER CD (provided with the kit) on the computer.

2. On the Home page, click System Initialization.

3. Click Import Kit at the bottom right side of the window. The Import Calibration or

Performance Kit dialog box opens.

4. Select Locate the CD in the appropriate drive and select the *.mpx folder and click

Open.

5. Select the *.lxl file and click Open.

6. Click OK.

Creating Calibration and Verification Reports

1. Open the Results page, then open the Reports tab.

2. In the Report drop-down list, select Calibration and Verification Reports.

3. In the Type drop-down list, select ALL, CAL, VER, or Fluidics.

4. Type a Start date and a Through date for the date range you want to view.

5. Click Generate to display the report.

6. Use the left or right Page arrows to navigate to the different report pages.

7. Click Print to print the report, or Save to save the report.

8. Click New Report to generate another report.

For In Vitro Diagnostic Use

Starting the MAGPIX System for the First Time

7