Magpix new system startup errors – Luminex xPONENT for MAGPIX IVD User Manual

Page 24

Advertising
background image

1. Adjusting the Sample Probe Height

2. Revive After Storage Routine (Luminex)

3. System Initialization

All system startup routines are run from the MAGPIX Off-Plate Reagent Block. For
detailed instructions on how to perform these operations, refer to the

Starting the

MAGPIX System for the First Time

section of this manual.

FIGURE 6.

MAGPIX Off-Plate Reagent Block

1

Reservoirs

2

Strip Wells

MAGPIX New System Startup Errors

Problem Summary

Resolution

Notes

Calibration Failure

(2x) Clean with 70% isopropanol
or ethanol alcohol (reservoir B)

(2x) Alcohol Flush with 70%
isopropanol or ethanol alcohol
(reservoir B)

(3x) Rinse (reservoir D)

(3x) Prime

Run Calibration Routine

Verify that reagent is freshly
vortexed. Use 6 drops of
reagent.

MAGPIX

®

User Quick Guide

18