Multichannel Systems MEA2100-System Manual User Manual

Page 19

Advertising
background image

Software

15

Selection of stimulation electrodes

Click the “Channels” tab to assign the programmed stimulation patterns to individual electrodes,
and to activate the blanking circuit BC, and to operate the “List Mode”.

To apply the stimulation pattern of a specific stimulation channel (Stim. 1, 2 or 3) to one or more
electrodes, select the desired stimulation channel (Stim. 1 to 3) and click the electrodes you want
to apply this paradigm to. To apply the stimulation pattern to all electrodes, click the button “All”
below the grid. Electrodes can be grounded or selected as regular recording electrodes (Not Stim.)
in the same way.

Electrodes selected for stimulation pattern 1 “Stim.1” are colored in blue, for stimulation pattern
2 “Stim. 2” are colored in red, or stimulation pattern 3 “Stim. 3" are colored in yellow. To cancel a
stimulation electrode, please use the radio button “Not Stim.” and click on the selected electrode.

The blanking feature to avoid stimulation artifacts can be activated or deactivated via the check
box “Blanking”.

For the MEA2100-32-System the “Channels” tab is adapted respectively. The function
“Dedicated Stimulation Electrodes” is not available.