Multichannel Systems MEA2100-System Manual User Manual

Page 44

Advertising
background image

MEA2100-System Manual

40

Setting up MC_Rack

Please refer to chapter "First Use of MC_Rack" and to the MC_Rack manual for more information.

Start the Data Acquisition

Click the “Start” button to start the data acquisition. No data is recorded. You see the raw data
streams of all channels. You may have to adjust the scaling of the y-axes until you can clearly see
the noise level. You should see the baseline with a maximum noise level of +/– 20 μV. Please note
that the system has a broad frequency range by default, so slow wave noise and drifting baselines
may occur. This can be solved by using virtual filters. See MC_Rack manual for details.

The display will be automatically adapted when using a MEA with 32 recording and 12 stimulation
electrodes, for example.

Advertising