2 taring – Nova-Tech WBZ..aM Series User Manual

Page 11

Advertising
background image

© Adam Equipment Company 2010

9

10.2 TARING

• Taring is used to eliminate the weight of a container so that only

the net weight is displayed. On the WBZ 15 scales the tare is
limited to 9.995kg. Weight above this amount will not be tared.

• Zero the scale by pressing [

Z

] if necessary. The zero

indicator will be on.

• Place an empty container on the platform. The weight of the

container will be displayed.

• Press [Tare] to zero the scale. The weight that was displayed is

stored as the tare value and that value is subtracted from the
display, leaving zero on the display. The Net indicator will be on.
As product is added only the weight of the product will be shown.

• When the product and container are removed the weight display

will show a negative value. If a different container is to be used,
remove this negative tare value by pressing the [Tare] once more.

Note: If the capacity of the scale is 6000g and a container weighing
2350g is used and tared, the scale can then be used to weigh the
material of up to 3650g.