2 without temperature probe (no atc) – Nova-Tech Ion 6+ User Manual

Page 17

Advertising
background image

Instruction Manual

pH 5+, pH 6+, Ion 6+

-14-

3.4.2

Without Temperature Probe (no ATC)

If no temperature probe is used, the meter compensates for pH response based on a
temperature value manually set by you or at 25.0

o

C (factory default).

1. Press

MODE key to enter into Temperature mode until “

o

C” shows in LCD.


2. Compare the displayed value to NIST certified thermometer or thermometer known

to be accurate (dipped into a constant temperature bath).

3. Press CAL key to enter temperature calibration mode. The LCD shows “CA

momentarily and displayed reading flashes. Note that this displayed value should
either be 25.0

o

C or last set temperature value.


4. Press

S and T key (for Ion 6) or INC key (for pH 5/6) until the displays shows the

desired temperature. You can set any value from 0 to 100

o

C.

5. To cancel or abort this operation, press CAL key. Note no new value will be stored

into its meter‘s non-volatile memory. To confirm calibration, press ENTER key. The
LCD displays “CO” momentarily, and the meter reverts to measurement mode.