Electrode care and maintenance, Troubleshooting – Nova-Tech Ion 6+ User Manual

Page 19

Advertising
background image

Instruction Manual

pH 5+, pH 6+, Ion 6+

-16-

5. ELECTRODE CARE AND MAINTENANCE

For best results, keep the ISE capped dry and pH/ORP electrode bulb wet. Store the
pH/ORP glass bulb with pH electrode storage solution. NEVER use deionised water for
storage. Wash electrodes with clean water after each use.

Your ISE or pH electrode is susceptible to contamination or dirt. Clean as needed using
mild detergent and warm water. Blot the probe gently with a soft tissue paper. Avoid
excessive drying of the glass membrane and avoid touching it with your fingers.
Recalibrate after cleaning.

6. TROUBLESHOOTING

Problem Cause

Solution

No display

Batteries not in place.

a) Insert batteries.

b) Re-insert batteries in correct polarity.

“---“ on
display

Ion 6+ does not have 2
point calibration

Perform 2 or 3 point ion calibration.

“LO” displays
in the LCD

Low battery

Replace batteries.

Unstable
reading

a) Electrode not deep

enough in sample

b) Dirty electrode.

c) Broken electrode

a) Place electrode deeper in sample.

b) Clean electrode and recalibrate.

c) Replace electrode.

“Er1” display

Buffer value out of
tolerance

Use new pH buffer solution and recalibrate. Ensure
correct pH buffer group was selected.

“Er2” display

Single point calibration

Perform at least 2 point calibration. (Ion 6+).

“Er3” display

ISE slope not within the
specified tolerance

Check ISE is in good working condition-refer to ISE
manual (Ion 6+).

“Er4” display

Any calibration points
not within 1 decade

Ensure any calibration points between each other
must be within 1 decade. (Ion 6+)

Not able to
calibrate

a) Display freezes

b) Faulty electrode

c) Inaccurate buffer

a) Release reading by pressing HOLD.

b) Replace electrode.

c) Replace expired buffer solutions.