TELEDYNE Model 9110AH NO
X
Analyzer Operator Manual, 01620, Rev. F
2-13
Table 2-1: Final Test and Calibration Values
TEST Values
Observed
Value
Units
Nominal Range
Reference Section
RANGE
ppm
5-5000
5.3.4
NOISE
ppm
0.0 - 0.2
9.1.1, Table 9-1, 9.2.5
SAMP FLW
cc/min
300 50 (Default)
550 50 (Optional)
9.3.7, Table 9-1
OZONE FL
cc/min
250 15
9.3.6
PMT
mV
0-5000
9.3.8
AUTOZERO
mV
-10 to +50
4.1
HVPS
V
400 - 700 constant
9.3.8.5
DCPS
mV
2500 200
9.3.5
RCELL TEMP
o
C
50 2
9.3.8.2
BLOCK TEMP
o
C
50 2
9.3.4.1
BOX TEMP
o
C
8-48
9.3.4.1
PMT TEMP
o
C
7 1
9.3.8.4
CONV TEMP
o
C
700 10 (Std)
315 5 (Moly)
9.3.4.1
RCEL PRES
IN-Hg-A
2 - 10 constant
9.3.7
SAMP PRES
IN-Hg-A
25 - 30 constant
9.3.7
Electric Test &Optic Test
Electric Test
PMT Volts
mV
2000 200
9.1.3.2
NO Conc
ppm
250 25
9.1.3.2
NO
x
Conc
ppm
250 25
9.1.3.2
OPTIC TEST
PMT Volts
mV
100 20
9.1.3.3
NO Conc
ppm
12.5 2
9.1.3.3
NO
x
Conc
ppm
12.5 2
9.1.3.3
(table continued)