6 diagnostic commands and messages – Teledyne 9110AH - Nitrogen oxides analyzer User Manual

Page 78

Advertising
background image

TELEDYNE Model 9110AH NO

X

Analyzer Operator Manual, 01620, Rev. F

5-36

Whenever the analyzer starts or finishes a calibration, it issues a status report to the RS-232
interface. If the RS-232 interface is in the normal mode, these reports will be sent. Otherwise,
they will be discarded. The format of these messages is:

C DDD:HH:MM IIII CALIBRATION STATUS MESSAGE

An example of an actual sequence of calibration status messages is:

C DDD:HH:MM IIII START MULTI-POINT CALIBRATION

C DDD:HH:MM IIII FINISH MULTI-POINT CALIBRATION

There are several methods of both checking the calibration and calibrating the M9110AH, these
are discussed in Section 7. C LIST lists the calibration commands available. The C command
executes a calibration command, which may be one of the following:

Table 5-18: RS-232 Calibration Commands

Command

Description

C LIST

List calibration commands

C ZERO

Enter zero calibration mode

C SPAN

Enter span calibration mode

C COMPUTE ZERO

Do a zero calibration

C COMPUTE SPAN

Do a span calibration

C EXITZ

Exit zero calibration

C EXITS

Exit span calibration

C EXIT

Exit zero or span calibration

5.7.6 DIAGNOSTIC Commands and Messages

When Diagnostic mode is entered from the RS-232 port, the diagnostic mode issues additional
status messages to indicate which diagnostic test is currently selected. Examples of Diagnostic
mode messages are:

D DDD:HH:MM IIII ZERO VALVE=ON

D DDD:HH:MM IIII ENTER DIAGNOSTIC MODE

D DDD:HH:MM IIII EXIT DIAGNOSTIC MODE