Teledyne 9110AH - Nitrogen oxides analyzer User Manual
Page 27

TELEDYNE Model 9110AH NO
X
Analyzer Operator Manual, 01620, Rev. F
2-14
Table 2-1: Final Test and Calibration Values (Continued)
Parameter
Observed
Value
Units
Nominal Range
Reference Section
NO Span Conc
ppm
0.5 - 5000
Table 7-3
NO
x
Span Conc
ppm
0.5 - 5000
Table 7-3
NO Slope
-
1.0 0.3
7.1, 7.9
NO
x
Slope
-
1.0 0.3
7.1
NO Offset
mV
25
7.1, 7.2
NO
x
Offset
mV
25
7.1, 7.2
Conv Efficiency
%
0.75 - 1.10
7.10, 5.2.2.6
Noise at Zero
ppm
0.0 - 0.2
Table 9-1
Noise At Span
ppm
0.1 - 0.5
Table 9-1
Measured Flows
Sample Flow
cc/min
50 20
9.3.7, Figure 9-8
Bypass Flow
cc/min
250 (Std)
500 (Optional)
Figure 8-4
Ozone Flow
cc/min
250 15
9.3.7, Figure 9-8
Factory Installed Options
Option Installed
Power Voltage/Frequency
Rack Mount, w/ Slides
Rack Mount, w/ Ears Only
Rack Mount, External Pump w/o Slides
Stainless Zero/Span Valves
4-20 mA Current Loop Output, Isolated
Bypass flow 500 cc/min
Molybdenum Converter
Desiccant Canister - O3 generator
PROM #
Serial #
Date
Technician